Air Leak Test Fixture
Design a test fixture to detect a leak on the order of 0.05ml over 10 minutes.
Problem
-
Detect a leak in a product equivalent to a leak that would cause failure during a 5-hour treatment.
Challenge
-
The off-the-shelf leak tester had an internal leak rate greater than the leak being tested.
Solution
-
Implement a manometer-type fixture that displays a drop in water level equivalent to the amount of air leaked from the product.
-
The solution relies only on a volumetric leak. To determine a pressure leak rather than a volumetric leak, the product volume would need to be known to a high degree of accuracy for each product; which is not possible given the variation from the assembly process.
![](https://static.wixstatic.com/media/785403_ac95a49c01e94240a3f61d77d936eb8d~mv2.jpg/v1/crop/x_444,y_202,w_3232,h_2822/fill/w_449,h_392,al_c,q_80,usm_0.66_1.00_0.01,enc_avif,quality_auto/785403_ac95a49c01e94240a3f61d77d936eb8d~mv2.jpg)
![](https://static.wixstatic.com/media/785403_3910de0c701d431dbf0b4231cf9289e5~mv2.jpg/v1/fill/w_600,h_450,al_c,q_80,usm_0.66_1.00_0.01,enc_avif,quality_auto/785403_3910de0c701d431dbf0b4231cf9289e5~mv2.jpg)
Product Failure Mode
A failure occurs during a 5-hour treatment if the leak is large enough to cause the membrane to occlude the orifice prior to the completion of the treatment.
![](https://static.wixstatic.com/media/785403_d0b0e8462b4249628d25f7273830742f~mv2.jpg/v1/fill/w_586,h_475,al_c,q_80,enc_avif,quality_auto/785403_d0b0e8462b4249628d25f7273830742f~mv2.jpg)